Coupling-Decoupling Networks Calibration

Intertek's process follows the requirements of IEC 61000-4-6, the basic standard, with resulting data collection providing very granular resolution.

While this procedure may produce many data points, we will reduce the coupling factor data to a manageable value while maintaining an accuracy of 0.1 dB with respect to the full breath of data. Impedance data will be plotted and compared to requirements listed in IEC 61000-4-6.

The table shown below describes our scope of accreditation for coupling decoupling network calibrations:
 

Parameter/Equipment  Range  Best Uncertainty (±)  Comments 
CDN (50 to 150) Ω     IEC 61000-4-6
Coupling Factor 150 kHz to 230 MHz 0.49 dB  
Impedance 150 kHz to 230 MHz 0.38 Ohms  

to see how Intertek can help your organization with Coupling-Decoupling Networks Calibration.