Oxidative Induction Times (OIT) Testing by Differential Scanning Calorimetry

The OIT test characterizes the thermo-oxidative stability of polyolefins, particularly polyethylene.

The Oxidation Induction Times Test (OIT), as carried out in a differential scanning calorimeter (DSC), is used to predict thermo-oxidative performance of a material.

Samples are heated up under a nitrogen atmosphere, typically to 200C. Oxygen is then introduced to the sample cell, and the length of time before the onset of degradation - as seen by the initiation of an exothermic process in the DSC trace - is measured.  

OIT is a sensitive measure of the level of anti-oxidative additives within the polymer.

 

 
Need help or have a question? +1 800 967 5352
 

Need help or have a question?

+1 800 967 5352
EMEA UK:
+44 161 721 5247
Benelux:
+31 88 126 8888
France:
+33 (0) 385 99 12 80
APAC:
+65 6805 4800
Germany:
0800 5855888
+49 711 27311 152
Switzerland:
+41 61 686 4800
Mexico:
01 800 5468 3783
+52 55 5091 2150
Brazil:
+55 11 2322 8033
Australia:
+61 1300 046 837
India:
+91 (22) 4245 0204
UAE
+971 4 317 8777